VHS-6
Diffusion Pump
Instruction Manual
Varian Vacuum Technologies
121 Hartwell Avenue
Lexington, MA 02421-3133
(781) 861-7200
(800) 8-VARIAN
Manual No. 6999-01-022
Revision E
July 2000
Zobrazit stránku 0
1 2 3 4 5 6 ... 29 30

Shrnutí obsahu

Strany 1 - Diffusion Pump

VHS-6Diffusion PumpInstruction ManualVarian Vacuum Technologies121 Hartwell AvenueLexington, MA 02421-3133(781) 861-7200(800) 8-VARIANManual No. 6999-

Strany 2 - Sales and Service Offices

viii. Cold Coolant – Liquid nitrogen, a cryogenic liquid, is used in traps. If it is splashed on body tissuesor eyes, it can cause severe frostbite or

Strany 3

viiiWARNING_____________High voltages can kill. Always break the primary circuit to the powersupply before starting to work on the heater and/or its w

Strany 4

ixNOTE________Always dispose of used or dirty oil properly and in compliance with alllocal, state, and federal environmental laws and regulations.CAUT

Strany 5

xTable of ContentsSection PageSection I – Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 6

1-1SECTION IINSTALLATIONBefore unpacking and installing the VHS-6 Diffusion Pump, the user should thoroughly familiarizehimself with this instruction

Strany 7

Installation1-2Heater circuit resistance (ohms) 2200 watts120V 6.5208V 19.6240V 26.1Clearance (for heater removal) 6 inches minimumMaterialsBody, flan

Strany 8 - Prohibited Action Result

WARNINGDiffusion pump heaters operate at voltages high enough to kill throughelectrical shock. During installation, check the drawings and be sure toa

Strany 9

Installation1-41-2-5 Electrical Connectionsa The diffusion pump has been designed to operate at a specific voltage. The voltage isspecified on the lab

Strany 10

2-1switch. This should be done only after the root cause of a problem has been determined andthe appropriate corrective action taken.SECTION IICAUTION

Strany 11

Operation2-22-2 SHUTDOWN PROCEDUREaTurn off the power to the diffusion pump. Continue to back the diffusion pump with the appro-priate mechanical pump

Strany 12 - ________

AmericaVarian Vacuum Technologies121 Hartwell AvenueLexington, MA 02421, USATel: (781) 861 7200Fax: (781) 860 5437Toll Free: 1 (800) 882 7426Argenti

Strany 13

3-1c After isolating the backing pump, the diffusion pump may be vented to atmosphere.d If faster cooling is desired, the pump can be cooled using the

Strany 14 - INSTALLATION

Maintenance and Servicec Unbolt the inlet flange and foreline connections. Remove the pump from the system.dDrain the diffusion pump of all fluid.e Re

Strany 15

Maintenance and Service3-3Figure 3-1 Cold Cap Removal/Installatione Install the spring (1) attached by the screw, (2) and washers (3 and 4). Torque th

Strany 16

Maintenance and Service3-4b Insert the jet base, making sure the ejector is aligned with the foreline (a keyway in the jetbase will assure alignment).

Strany 17

4-1SECTION IVTROUBLESHOOTING4-1 LEAKAGEAnalysis of general operational experience with diffusion pumps indicates that certain locations aremore prone

Strany 18 - OPERATION

Troubleshooting4-2Table 4-1 TroubleshootingFault Probable Cause Corrective ActionPoor system pressure Leaks in system, virtual or real Locate and repa

Strany 19

5-1SECTION VREPLACEMENT PARTSThe following parts are available from Varian; please call the 800 number listed inside the front coverof this manual.Tab

Strany 21

Replacement Parts5-3

Strany 22

Replacement Parts5-4VHS-6ASA ISO ConFlatin mm in mm in mmA Height21.92 557 21.92 557 21.92 557B Centerline to Centerline13.38 340 13.38 340 13.38 340C

Strany 23

WARRANTYProducts manufactured by Seller are warranted against defects in materials and workmanship fortwelve (12) months from date of shipment thereof

Strany 24 - TROUBLESHOOTING

Replacement Parts5-5Note: Speed curves were generated according to AVS Standard 4.11 Torr - 133 PaVHS-6 Speed and Throughput Curves

Strany 25 - Table 4-1 Troubleshooting

iOPERATION AND MAINTENANCE OF THIS EQUIPMENT INVOLVES SERIOUS RISK. IT IS THERESPONSIBILITY OF THE USER TO MAINTAIN SAFE OPERATING CONDITIONS AT ALL T

Strany 26 - Table 5-1 Troubleshooting

iiSAFE INCORPORATION OF VARIAN DIFFUSION PUMPS INTO VACUUM SYSTEMS IS THERESPONSIBILITY OF THE SYSTEMS DESIGNER. TAKE APPROPRIATE ACTION THROUGHREDUND

Strany 27 - Replacement Parts

iiiTHE INSTALLATION, OPERATION, AND/OR SERVICING OR DIFFUSION PUMPS INVOLVES ONEOR MORE OF THE FOLLOWING HAZARDS, ANY ONE OF WHICH, IN THE ABSENCE OF

Strany 28

ivHazard Suggested Corrective ActionExplosion Integrate pressure relief valves in all systems using pumps 10 inches or larger in diameter.Do not use o

Strany 29

vProhibited Action Explosion-Causing ConditionOpen drain or fill plug while pump is Overpressure plus strong oxidizerunder vacuum, especially when it

Strany 30 - 1 Torr - 133 Pa

vid. Poisonous and/or Corrosive Compounds – When pumping poisonous, reactive, and/or corrosivegas, vapors, or chemicals, even proper operation and reg